A signal model for fat-suppressed T1-mapping and dynamic contrast-enhanced MRI with interrupted spoiled gradient-echo readout

  • Myrte Wennen*
  • , Wilhelm Stehling
  • , J. Tim Marcus
  • , Joost P. A. Kuijer
  • , Cristina Lavini
  • , Leo M. A. Heunks
  • , Gustav J. Strijkers
  • , Bram F. Coolen
  • , Aart J. Nederveen
  • , Oliver J. Gurney-Champion
  • *Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

1 Citation (Scopus)
75 Downloads (Pure)

Abstract

The conventional gradient-echo steady-state signal model is the basis of various spoiled gradient-echo (SPGR) based quantitative MRI models, including variable flip angle (VFA) MRI and dynamic contrast-enhanced MRI (DCE). However, including preparation pulses, such as fat suppression or saturation bands, disrupts the steady-state and leads to a bias in T 1 and DCE parameter estimates. This work introduces a signal model that improves the accuracy of VFA T 1-mapping and DCE for interrupted spoiled gradient-echo (I-SPGR) acquisitions. The proposed model was applied to a VFA T 1-mapping I-SPGR sequence in the Gold Standard T 1-phantom (3 T), in the brain of four healthy volunteers (3 T), and to an abdominal DCE examination (1.5 T). T 1-values obtained with the proposed and conventional model were compared to reference T 1-values. Bland–Altman analysis (phantom) and analysis of variance (in vivo) were used to test whether bias from both methods was significantly different (p = 0.05). The proposed model outperformed the conventional model by decreasing the bias in the phantom with respect to the phantom reference values (mean bias -2 vs. -35% at 3 T) and in vivo with respect to the conventional SPGR (-6 vs. -37% bias in T 1, p < 0.01). The proposed signal model estimated approximately 48% (depending on baseline T 1) higher contrast concentrations in vivo, which resulted in decreased DCE pharmacokinetic parameter estimates of up to 35%. The proposed signal model improves the accuracy of quantitative parameter estimation from disrupted steady-state I-SPGR sequences. It therefore provides a flexible method for applying fat suppression, saturation bands, and other preparation pulses in VFA T 1-mapping and DCE.

Original languageEnglish
Article numbere5289
Number of pages15
JournalNMR in Biomedicine
Volume38
Issue number1
DOIs
Publication statusPublished - Jan 2025

Bibliographical note

Publisher Copyright:
© 2024 The Author(s). NMR in Biomedicine published by John Wiley & Sons Ltd.