Accuracy in prediction of catheter rotation in IVUS with feature-based optical flow - A phantom study

Mike Danilouchkine, Frits Mastik, Ton van der Steen

Research output: Contribution to journalArticleAcademicpeer-review

17 Citations (Scopus)

Abstract

The quantitative assessment of and compensation for catheter rotation in intravascular ultrasound images presents a fundamental problem for noninvasive characterization of the mechanical properties of the coronary arteries. A method based on the scale-space optical flow algorithm with a feature-based weighting scheme is proposed to account for the aforementioned artifact. The computed vector field, describing the misalignment between two consecutive frames, allows the quantitative assessment of the amount of vessel wall tissue motion, which is directly related to the catheter rotation. Algorithm accuracy and robustness were demonstrated on two tissue-mimicking phantoms, subjected to controlled amount of angular deviation. The proposed method shows a great reliability in the prediction of catheter rotational motion up to 4 degrees.
Original languageUndefined/Unknown
Pages (from-to)356-365
Number of pages10
JournalIEEE Transactions on Information Technology in Biomedicine
Volume12
Issue number3
DOIs
Publication statusPublished - 2008