Abstract
Aims: To validate a new two dimensional (2-D) bifurcation quantitative coronary angiography (QCA) software. Methods and results: In the latest edition of the Cardiovascular Angiography Analysis System (CAAS 5.9; Pie Medical Imaging, Maastricht, The Netherlands) video-densitometric information is dynamically integrated into the edge-detection algorithm of 11- and 6-segment models to reduce overestimation of small diameters. Furthermore, automatic reference obstruction analysis was optimised. Values of the minimal lumen diameter (MLD), reference vessel diameter (RVD), percent diameter stenosis (DS) and Conclusions: MLD quantification is improved for true MLD <0.7 mm, resulting in highly accurate and precise diameter measurements over the entire range of phantom diameters. Automatic reference obstruction analysis provides highly accurate, precise and reproducible RVD and DS measurements.
| Original language | Undefined/Unknown |
|---|---|
| Pages (from-to) | 1326-1335 |
| Number of pages | 10 |
| Journal | EuroIntervention |
| Volume | 7 |
| Issue number | 11 |
| DOIs | |
| Publication status | Published - 2012 |
Research programs
- EMC COEUR-09
Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver