Afterloader integrated EMT enables improved dwell position model definition and quality assurance in Venezia gynaecological brachytherapy applicators

Research output: Contribution to journal › Article › Academic › peer-review

3 Citations (Scopus)
37 Downloads (Pure)

Abstract

Objective:

In brachytherapy for gynecological cancers using intracavitary applicators, implant reconstruction is commonly performed using applicator libraries. These libraries contain applicator geometry models as well as dwell position (DP) models defined in respect to the applicator geometry. In this study, we investigate whether an afterloader integrated electromagnetic tracking (EMT) system can be utilized for DP model definition and quality assurance in such applicators.

Approach:

DPs in four sets of two configurations of the Elekta Venezia Advanced Gynaecological Applicator (22 mm ovoids/40 mm intrauterine (IU) and 26 mm ovoids/70 mm IU) were measured using an afterloader integrated EMT system. Measurements were evaluated for reproducibility and compared against manufacturer-specified (MS) DPs and a computed tomography (CT)-corrected DP model.

Main Results:

Excellent EMT measurement reproducibility was observed, with values of ⩽0.2 mm for both configurations. The overall reproducibility, including applicator geometry reproducibility, was ⩽0.4 mm for both configurations. Significant discrepancies from the manufacturer's DP model were observed, with a mean ± sd deviation of 1.13 ± 0.66 mm (22/40) and 1.37 ± 0.63 (26/70), particularly in the IU channel, where MS DPs were not experimentally defined. Discrepancies were reduced to 0.89 ± 0.41 mm (22/40) and 0.81 ± 0.33 mm (26/70) when the CT-corrected DP model was used as baseline, highlighting the need for experimentally defined DP models. The overall uncertainty of single measurements was below the clinically acceptable 2 mm limit.

Significance:

This study confirms that afterloader integrated EMT can accurately reconstruct source paths in gynecological brachytherapy applicators and supports its incorporation into clinical workflows for improved quality assurance and treatment precision. The importance of EMT for quality assurance was highlighted by measured deviations from manufacturer's DP model in a clinical relevant part of the IU channel.
Original languageEnglish
Article number125003
JournalPhysics in Medicine and Biology
Volume70
Issue number12
DOIs
Publication statusPublished - 22 Jun 2025

Bibliographical note

Publisher Copyright:
© 2025 The Author(s). Published on behalf of Institute of Physics and Engineering in Medicine by IOP Publishing Ltd.

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being

Fingerprint

Dive into the research topics of 'Afterloader integrated EMT enables improved dwell position model definition and quality assurance in Venezia gynaecological brachytherapy applicators'. Together they form a unique fingerprint.

Cite this