Abstract

Purpose: To describe the imaging characteristics of the Nellix Endovascular Aneurysm Sealing (EVAS) System on serial computed tomography (CT) surveillance. Methods: Sixty-eight patients undergoing EVAS were enrolled in a surveillance protocol that included CT scans prior to hospital discharge and at 3, 6, and 9 months postoperatively. Images were analyzed for the presence of gas within the endobag, endoleak, and for maximum radiodensity measured in Hounsfield units (HU) within the uppermost, middle, and lowermost regions of each endobag. Results: Gas was seen within the endobags of all 68 EVAS repairs at the first postoperative CT compared with 2 (5.6%) of 36 undergoing the 3-month scan. The endobags appeared radiodense during initial imaging, and the median (interquartile range) radiodensity of the Nellix polymer decreased from 158.3 HU (149.5; 169.5) at the postoperative CT to 81.0 HU (74.0; 88.0) at 3 months, excluding 3 cases in which contrast pre-fill was utilized. Type I endoleak was seen at the periphery of the aneurysm sac or in the cleft between the endobags, with a substantially different appearance to endoleak after endovascular aneurysm repair. Conclusion: The evolution of CT appearances after EVAS was characteristic and predictable. The device endobags were initially radiodense, which may impact the detection of endoleak within 3 months of EVAS. Endoleaks after EVAS were seen in a different anatomical area to endoleaks after conventional stent-graft repair.

Original languageEnglish
Pages (from-to)297-302
Number of pages6
JournalJournal of Endovascular Therapy
Volume22
Issue number3
DOIs
Publication statusPublished - 1 Jun 2015
Externally publishedYes

Bibliographical note

Publisher Copyright:
© The Author(s) 2015.

Fingerprint

Dive into the research topics of 'Appearance of the Nellix endovascular aneurysm sealing system on computed tomography: Implications for postoperative imaging surveillance'. Together they form a unique fingerprint.

Cite this