Blood-flow Volume Estimation by a 2-D Sparse Array

  • Claudio Giangrossi*
  • , Alessandro Ramalli
  • , Francesco Guidi
  • , Emile Noothout
  • , Luxi Wei
  • , Hendrik J. Vos
  • , Piero Tortoli
  • *Corresponding author for this work
  • University of Florence
  • Delft University of Technology
  • Erasmus University Rotterdam

Research output: Contribution to journalArticleAcademicpeer-review

3 Citations (Scopus)
57 Downloads (Pure)

Abstract

Objective: 

The assessment of blood-flow volume (BFV) is clinically relevant for the diagnosis and monitoring of cardiovascular dysfunctions and the prevention of subsequent secondary diseases. Non-invasive BFV measurement based on ultrasound methods are appealing for lower cost, real-time operation, and equipment portability. Recently, complex ultrasound research scanners with 1024 channels controlling the elements of a 2-D matrix array probe, have been demonstrated suitable for off-line accurate BFV estimates. In this work, a streamlined approach, using a 256-channel research scanner paired with a 256-element 2-D sparse spiral array, is proposed and validated. 

Methods: 

This setup allows for simultaneous scanning of the vessel's longitudinal and transverse sections through an interleaved transmission sequence. In real-time, the longitudinal scan is used to determine the flow direction, while the transverse scan captures both the dynamic cross-sectional area and the local velocities by high frame rate color flow mapping. 

Results: 

Flow phantom experiments under steady and pulsatile flow conditions were conducted to assess the performance by comparing the measurements with the outputs of a reference flow sensor. The proposed method provided accurate and precise BFV values for both flow conditions, with mean percentage error and standard deviation always lower than 9.4% and 2.8%, respectively. Furthermore, preliminary in vivo experiments have produced results consistent with those reported in the literature. 

Conclusion: 

The proposed method based on the use of a sparse array has permitted accurate and precise phantom BFV measurements and has been shown suitable for real-time arterial BFV measurements.

Original languageEnglish
Pages (from-to)1580-1588
Number of pages9
JournalUltrasound in Medicine and Biology
Volume51
Issue number9
DOIs
Publication statusPublished - Sept 2025

Bibliographical note

Publisher Copyright:
© 2025 The Authors

Fingerprint

Dive into the research topics of 'Blood-flow Volume Estimation by a 2-D Sparse Array'. Together they form a unique fingerprint.

Cite this