Carotid artery assessment in dual-source photon-counting CT: impact of low-energy virtual monoenergetic imaging on image quality, vascular contrast and diagnostic assessability

  • Christian Booz
  • , Giuseppe M. Bucolo
  • , Tommaso D’Angelo*
  • , Silvio Mazziotti
  • , Ludovica R.M. Lanzafame
  • , Ibrahim Yel
  • , Leona S. Alizadeh
  • , Leon D. Gruenewald
  • , Vitali Koch
  • , Simon S. Martin
  • , Mirela Dimitrova
  • , Aynur Goekduman
  • , Thomas J. Vogl
  • , Hanns L. Kaatsch
  • , Daniel Overhoff
  • , Stephan Waldeck
  • *Corresponding author for this work
  • University Hospital Frankfurt am Main
  • University of Messina
  • University Hospital “Policlinico G. Martino”
  • Bundeswehr Central Hospital Koblenz
  • Heidelberg University 
  • Johannes Gutenberg University Mainz

Research output: Contribution to journal › Article › Academic › peer-review

15 Citations (Scopus)
53 Downloads (Pure)

Abstract

Purpose: 

Preliminary dual-energy CT studies have shown that low-energy virtual monoenergetic (VMI) + reconstructions can provide superior image quality compared to standard 120 kV CTA series. The purpose of this study is to evaluate the impact of low-energy VMI reconstructions on quantitative and qualitative image quality, vascular contrast, and diagnostic assessability of the carotid artery in patients undergoing photon-counting CTA examinations. 

Materials and methods: 

A total of 122 patients (67 male) who had undergone dual-source photon-counting CTA scans of the carotid artery were retrospectively analyzed in this study. Standard 120 kV CT images and low-keV VMI series from 40 to 100 keV with an interval of 15 keV were reconstructed. Quantitative analyses included the evaluation of vascular CT numbers, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR). CT number measurements were performed in the common, external, and internal carotid arteries. Qualitative analyses were performed by three board-certified radiologists independently using five-point scales to evaluate image quality, vascular contrast, and diagnostic assessability of the carotid artery. 

Results: 

Mean attenuation, CNR and SNR values were highest in 40 keV VMI reconstructions (HU, 1362.32 ± 457.81; CNR, 33.19 ± 12.86; SNR, 34.37 ± 12.89) followed by 55-keV VMI reconstructions (HU, 736.94 ± 150.09; CNR, 24.49 ± 7.11; SNR, 26.25 ± 7.34); all three mean values at these keV levels were significantly higher compared with the remaining VMI series and standard 120 kV CT series (HU, 154.43 ± 23.69; CNR, 16.34 ± 5.47; SNR, 24.44 ± 7.14) (p < 0.0001). The qualitative analysis showed the highest rating scores for 55 keV VMI reconstructions followed by 40 keV and 70 keV VMI series with a significant difference compared to standard 120 kV CT images series regarding image quality, vascular contrast, and diagnostic assessability of the carotid artery (all comparisons, p < 0.01). 

Conclusions: 

Low-keV VMI reconstructions at a level of 40–55 keV significantly improve image quality, vascular contrast, and the diagnostic assessability of the carotid artery compared with standard CT series in photon-counting CTA.

Original languageEnglish
Pages (from-to)1633-1643
Number of pages11
JournalRadiologia Medica
Volume129
Issue number11
DOIs
Publication statusPublished - Nov 2024