Abstract
Objectives: To date, over 200 000 cochlear implants (CIs) have been implanted worldwide and the incidence is still increasing. We present a case of traumatic displacement of CI magnet to raise awareness about this complication and to highlight the need for vigilance during surgery as well as for proper counseling. Methods: The clinical presentation of a 1.5-year-old boy with a traumatic displacement of a CI magnet was presented and the literature was reviewed for this rare complication. Results: After minor head injury, the sound processor could no longer connect to the CI. X-ray imaging demonstrated displacement of the CI magnet. During revision surgery, the magnet was replaced by a new magnet in the silicon holding cap. Intraoperative impedance measurements were normal and the CI was successfully activated 4 weeks postoperatively. Conclusions: Clinicians and patients should be aware of the risk of displacement of the CI magnet after (minor) head injury. Young boys tend to have a higher risk for this complication.
Original language | Undefined/Unknown |
---|---|
Pages (from-to) | 229-231 |
Number of pages | 3 |
Journal | Annals of Otology Rhinology and Laryngology |
Volume | 123 |
Issue number | 4 |
DOIs | |
Publication status | Published - 2014 |
Research programs