Feasibility of a Novel Augmented 6-Minute Incremental Step Test: A Simplified Cardiorespiratory Fitness Assessment Tool

Jeroen Molinger, Veraprapas Kittipibul, J. Matthew Gray, Vishal N. Rao, Stratton Barth, Ashley Swavely, Brian Coyne, Aubrie Coburn, Jan Bakker, Paul E. Wischmeyer, Cynthia L. Green, David MacLeod, Manesh Patel, Marat Fudim*

*Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

1 Citation (Scopus)
35 Downloads (Pure)

Abstract

Background: 

The cardiopulmonary exercise test (CPET) is considered a gold standard in assessing cardiorespiratory fitness (CRF) but has limited accessibility due to competency requirements and cost. Incorporating portable sensor devices into a simple bedside test of CRF could improve diagnostic and prognostic value. 

Objectives: 

The authors sought to evaluate the association of an augmented 6-minute incremental step test (6MIST) with standard CPET. 

Methods: 

We enrolled patients undergoing clinically indicated supine cycle ergometry CPET with invasive hemodynamics (iCPET) for the same-day 6MIST. CRF-related variables were simultaneously recorded using a signal morphology-based impedance cardiograph (PhysioFlow Enduro) and a portable metabolic analyzer (VO2 Master Pro) during incremental pace stationary stepping. The correlation between CPET and hemodynamic parameters from both tests was assessed using the intraclass correlation coefficient (ICC). 

Results: 

Fifteen patients (mean age 60 ± 14 years, 40% female, 27% Black) were included. All patients who agreed to undergo 6MIST completed the study without any test-related adverse events. We observed good to excellent correlation between iCPET- and 6MIST-measured CPET parameters: peak heart rate (ICC = 0.60; 95% CI: 0.15-0.85), absolute peak O2 consumption (VO2) (ICC = 0.77; 95% CI: 0.44-0.92), relative peak VO2 (ICC = 0.64; 95% CI: 0.20-0.86), maximum ventilation (ICC = 0.59; 95% CI: 0.13-0.84), O2 pulse (ICC = 0.71; 95% CI: 0.33-0.89), and cardiorespiratory optimal point (ICC = 0.82; 95% CI: 0.52-0.94). No significant correlation was determined between iCPET and 6MIST in measuring cardiac index at rest (ICC = 0.19; 95% CI: −0.34 to 0.63) or at peak exercise (ICC = 0.36; 95% CI: −0.17 to 0.73). 

Conclusions: 

We demonstrate the feasibility of a novel augmented 6MIST with wearable devices for simultaneous CPET and hemodynamic assessment. 6MIST-measured CPET parameters were strongly correlated with the iCPET-derived measurements. Additional studies are needed to confirm the validity of the 6MIST compared to standard upright CPET.

Original languageEnglish
Article number101079
JournalJACC: Advances
Volume3
Issue number8
DOIs
Publication statusPublished - Aug 2024

Bibliographical note

Publisher Copyright:
© 2024 The Authors