Geometric accuracy of a novel gimbals based radiation therapy tumor tracking system

T Depuydt, D Verellen, O Haas, T Gevaert, N Linthout, M Duchateau, K Tournel, T Reynders, K Leysen, Mischa Hoogeman, G Storme, Maria de Ridder

Research output: Contribution to journalArticleAcademicpeer-review

140 Citations (Scopus)

Abstract

Purpose: VERO is a novel platform for image guided stereotactic body radiotherapy. Orthogonal gimbals hold the linac-MLC assembly allowing real-time moving tumor tracking. This study determines the geometric accuracy of the tracking. Materials and methods: To determine the tracking error, an 1D moving phantom produced sinusoidal motion with frequencies up to 30 breaths per minute (bpm). Tumor trajectories of patients were reproduced using a 2D robot and pursued with the gimbals tracking system prototype. Using the moving beam light field and a digital-camera-based detection unit tracking errors, system lag and equivalence of pan/tilt performance were measured. Results: The system lag was 47.7 ms for panning and 47.6 ms for tilting. Applying system lag compensation, sinusoidal motion tracking was accurate, with a tracking error 90% percentile E(90%) < 0.82 mm and similar performance for pan/tilt. Systematic tracking errors were below 0.14 mm. The 2D tumor trajectories were tracked with an average E(90%) of 0.54 mm, and tracking error standard deviations of 0.20 mm for pan and 0.22 mm for tilt. Conclusions: In terms of dynamic behavior, the gimbaled linac of the VERO system showed to be an excellent approach for providing accurate real-time tumor tracking in radiation therapy. (C) 2011 Elsevier Ireland Ltd All rights reserved. Radiotherapy and Oncology 98 (2011) 365-372
Original languageUndefined/Unknown
Pages (from-to)365-372
Number of pages8
JournalRadiotherapy and Oncology
Volume98
Issue number3
DOIs
Publication statusPublished - 2011

Cite this