Improved understanding of stent malapposition using virtual bench testing

Peter Mortier*, Heleen M.M. van Beusekom, Matthieu De Beule, Ilona Krabbendam-Peters, Benjamin Van Der Smissen, Gianluca De Santis, Jurgen M. Ligthart, Benedict Verhegghe, Wim J. van der Giessen

*Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

6 Citations (Scopus)

Abstract

Intravascular imaging techniques such as optical coherence tomography (OCT) and intravascular ultrasound (IVUS) are often used to assess strut apposition, but only provide limited insight into the three-dimensional appositioning behaviour of stents. Recently, a new approach has been introduced to study the phenomenon of incomplete stent apposition (ISA) based on finite element simulations. In this study, we employed this virtual strut apposition assessment technique in the setting of coronary bifurcation stenting and compared simulated strut-artery distances of two stent designs with actual measurements based on OCT imaging using a silicone model. Stenting of the main branch leads to malapposed struts in the proximal part and the average strut-artery distance in that region for the Integrity stent is 126 μm based on the simulation and 117±14 μm based on the OCT analysis. For the Multi-Link 8 stent, this average distance is 150 μm and 174±7 μm for the simulation and the in vitro OCT measurements respectively. In conclusion, the virtual assessment of strut appositioning results in similar strut-artery distances when compared with measurements based on OCT-visualised in vitro stent deployments and could be used to optimise devices and procedures.

Original languageEnglish
Pages (from-to)106-109
Number of pages4
JournalInterventional Cardiology: Reviews, Research, Resources
Volume6
Issue number2
DOIs
Publication statusPublished - Aug 2011

Bibliographical note

Publisher Copyright: © TOUCH BRIEFINGS 2011.

The authors acknowledge Medtronic for providing a research grant for this study and for delivering all stent and balloon samples.

Fingerprint

Dive into the research topics of 'Improved understanding of stent malapposition using virtual bench testing'. Together they form a unique fingerprint.

Cite this