Abstract
Introduction: There is a growing use of locking compression plates in fracture surgery. The current study was undertaken to investigate the wound complication rates of locking versus non-locking plates in distal fibular fractures. Patients and methods: During a 6-year study period all consecutive, closed distal fibular fractures treated with either a locking or a non-locking plate were included and retrospectively analysed for complication related to the fibula. Results: A total of 165 patients received a one-third tubular plate and 40 patients were treated with a locking plate. The two groups were comparable with respect to patient characteristics (age, gender, smokers and diabetics), injury characteristics (affected side, fracture dislocations, number of fractured malleoli and classification) and operation characteristics (surgical delay and duration, use of a tourniquet and plate length). The wound complication rate was 5.5% in the conventional plati Conclusion: There is a significant increase in wound complications in distal fibular fractures treated with a locking compression plate. In light of the current study, we would caution against the application of the currently used locking compression plates in the treatment of distal fibular fractures. (C) 2011 Elsevier Ltd. All rights reserved.
| Original language | Undefined/Unknown |
|---|---|
| Pages (from-to) | 1125-1129 |
| Number of pages | 5 |
| Journal | Injury-International Journal of the Care of the Injured |
| Volume | 42 |
| Issue number | 10 |
| DOIs | |
| Publication status | Published - 2011 |
UN SDGs
This output contributes to the following UN Sustainable Development Goals (SDGs)
-
SDG 3 Good Health and Well-being
Research programs
- EMC MUSC-01-47-01
Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver