Abstract
Quantitative Magnetic Resonance Imaging (MRI) can enable early diagnosis of knee cartilage damage if imaging is performed during the application of load. Mechanical loading via ropes, pulleys and suspended weights can be obstructive and require adaptations to the patient table. In this paper, a new lightweight MRI-compatible elastic loading mechanism is introduced. The new device showed sufficient linearity (|α/β| = 0.42 ± 0.25), reproducibility (CoV = 5 ± 2%), and stability (CoV = 0.5 ± 0.1%). In vivo and ex vivo scans confirmed the ability of the device to exert sufficient force to study the knee cartilage under loading conditions, inducing up to a 29% decrease in T2 of the central medial cartilage. With this device mechanical loading can become more accessible for researchers and clinicians, thus facilitating the translational use of MRI biomarkers for the detection of cartilage deterioration.
Original language | English |
---|---|
Title of host publication | 43rd Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2021 |
Publisher | Institute of Electrical and Electronics Engineers Inc. |
Pages | 3721-3724 |
Number of pages | 4 |
ISBN (Electronic) | 9781728111797 |
DOIs | |
Publication status | Published - 2021 |
Event | 43rd Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2021 - Virtual, Online, Mexico Duration: 1 Nov 2021 → 5 Nov 2021 |
Publication series
Series | Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBS |
---|---|
ISSN | 1557-170X |
Conference
Conference | 43rd Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2021 |
---|---|
Country/Territory | Mexico |
City | Virtual, Online |
Period | 1/11/21 → 5/11/21 |
Bibliographical note
Funding Information:This study was approved by the Human Research Ethics Committee of the TU Delft
Publisher Copyright:
© 2021 IEEE.