Skip to main navigation Skip to search Skip to main content

Performance of time delay estimation methods for small time shifts in ultrasonic signals

  • Chris L. de Korte*
  • , Ton van der Steen
  • , B.H.J. Dijkman
  • , Charles T. Lancee
  • *Corresponding author for this work
  • Interuniversity Cardiology Institute of the Netherlands
  • Thoraxcentrum (Rotterdam)
  • Erasmus University Medical Centre

Research output: Contribution to journalArticleAcademicpeer-review

35 Citations (Scopus)

Abstract

In this study several time delay estimation (TDE) methods were investigated for estimation of time shifts of less than 10 ns at a frequency of 30 MHz. Using simulated and experimental echosignals we investigated the performance of five methods: two phase related methods (phase shift and phase difference method); two correlation methods (cross-correlation and correlation interpolation method); and a demodulation method. The results showed that the correlation interpolation method is by far the most accurate for all time delays. With this method, estimation errors of about 200 ps are achievable with an signal-to-noise ratio (SNR) of 40 dB (f0 = 30 MHz, bandwidth = 20 MHz) for time shifts of up to 10 ns.
Original languageEnglish
Pages (from-to)263-274
Number of pages12
JournalUltrasonics
Volume35
Issue number4
DOIs
Publication statusPublished - Jun 1997

Research programs

  • EMC 05-01-43-06-02

Fingerprint

Dive into the research topics of 'Performance of time delay estimation methods for small time shifts in ultrasonic signals'. Together they form a unique fingerprint.

Cite this