Abstract
The air-Q Intubating Laryngeal Airway (ILA) is a newly introduced extraglottic airway device. In this pilot study, we evaluated its use as a routine airway device during positive pressure ventilation. Ease of endotracheal intubation through the device was also assessed. Fifty-nine ASA I and II patients undergoing elective surgery received an air-Q ILA and an endotracheal tube where indicated. Insertion, ventilation and intubation characteristics were noted, as well as throat morbidity and occurrence of adverse events. An air-Q ILA was successfully inserted in 100% of patients. Mean leak pressure was 19±5 cmH2O. Endotracheal intubation was indicated in 19 patients and successful in 58% on the first attempt and 74% in total. Ten percent of the study patients were noted to have dysphagia. One patient was diagnosed with bilateral lingual nerve injury but made a complete recovery in four weeks. The air-Q ILA is an adequate extraglottic airway device in terms of insertion and ventilation. However, the proposed advantage of ease of endotracheal intubation requires further investigation.
| Original language | English |
|---|---|
| Pages (from-to) | 346-348 |
| Number of pages | 3 |
| Journal | Anaesthesia and Intensive Care |
| Volume | 38 |
| Issue number | 2 |
| Early online date | 1 Mar 2010 |
| DOIs | |
| Publication status | Published - 2010 |