Skip to main navigation Skip to search Skip to main content

Procedural Performance of Ultrathin, Biodegradable Polymer-Coated Stents Versus Durable Polymer-Coated Stents Based on Intracoronary Imaging

Research output: Contribution to journalArticleAcademicpeer-review

187 Downloads (Pure)

Abstract

OBJECTIVE: Thinner stent struts might lead to a higher risk of recoil and subsequently a smaller minimal stent area (MSA), which is known to be the strongest predictor of stent failure. We compared procedural performance between an ultrathin-strut biodegradable-polymer sirolimus-eluting stent (BP-SES) and a durable-polymer zotarolimus-eluting stent (DP-ZES) using intracoronary imaging.

METHODS: A consecutive cohort of patients underwent percutaneous coronary intervention (PCI) with either BP-SES or DP-ZES in a pseudorandomized fashion between July 2018 and October 2019. In the present subanalysis, we included cases in which post-PCI imaging with intravascular ultrasound (IVUS) or optical coherence tomography (OCT) was performed. The primary endpoint of the study was MSA. Secondary endpoints included percentage stent expansion and presence of residual edge disease, malapposition, tissue protrusion, submedial edge dissections, or edge hematoma.

RESULTS: A total of 141 treated lesions (78 BP-SES and 63 DP-ZES) in 127 patients were analyzed. Median age was 69.3 years (interquartile range [IQR], 57.3-75.6) and 74.0% of patients were male. All baseline and procedural characteristics were comparable between both groups. Median MSA was 5.80 mm² (IQR, 4.40-7.24) for BP-SES and 6.35 mm² (IQR, 4.76-8.31) for DP-ZES (P=.15). No significant differences in stent expansion, residual edge disease and presence of malapposition, tissue protrusion, submedial edge dissections, or edge hematomas were found. Stent diameter and stent length were found to be independent predictors of MSA.

CONCLUSIONS: No significant differences in MSA were found between lesions treated with BP-SES vs DP-ZES. BP-SES and DP-ZES were comparable in terms of procedural performance.

Original languageEnglish
Pages (from-to)E811-E819
JournalThe Journal of invasive cardiology
Volume34
Issue number11
Publication statusPublished - Nov 2022

Bibliographical note

Publisher Copyright:
© 2022 HMP Global.

Fingerprint

Dive into the research topics of 'Procedural Performance of Ultrathin, Biodegradable Polymer-Coated Stents Versus Durable Polymer-Coated Stents Based on Intracoronary Imaging'. Together they form a unique fingerprint.

Cite this