Roll and pitch set-up errors during volumetric modulated arc delivery: Can adapting gantry and collimator angles compensate?

  • VU University Medical Center

Research output: Contribution to journalArticleAcademicpeer-review

6 Citations (Scopus)

Abstract

Purpose: The purpose of this work was to investigate whether adapting gantry and collimator angles can compensate for roll and pitch setup errors during volumetric modulated arc therapy (VMAT) delivery. Methods: Previously delivered clinical plans for locally advanced head-and-neck (H&N) cancer (n = 5), localized prostate cancer (n = 2), and whole brain with simultaneous integrated boost to 5 metastases (WB + 5M, n = 1) were used for this study. Known rigid rotations were introduced in the planning CT scans. To compensate for these, in-house software was used to adapt gantry and collimator angles in the plan. Doses to planning target volumes (PTV) and critical organs at risk (OAR) were calculated with and without compensation and compared with the original clinical plan. Measurements in the sagittal plane in a polystyrene phantom using radiochromic film were compared by gamma (γ) evaluation for 2 H&N cancer patients. Results: For H&N plans, the introduction of 2°-roll and 3°-pitch rotations reduced mean PTV coverage from 98.7 to 96.3 %. This improved to 98.1 % with gantry and collimator compensation. For prostate plans respective figures were 98.4, 97.5, and 98.4 %. For WB + 5M, compensation worked less well, especially for smaller volumes and volumes farther from the isocenter. Mean comparative γ evaluation (3 %, 1 mm) between original and pitched plans resulted in 86 % γ < 1. The corrected plan restored the mean comparison to 96 % γ < 1. Conclusion: Preliminary data suggest that adapting gantry and collimator angles is a promising way to correct roll and pitch set-up errors of < 3° during VMAT for H&N and prostate cancer.

Original languageEnglish
Pages (from-to)272-280
Number of pages9
JournalStrahlentherapie und Onkologie
Volume191
Issue number3
Early online date28 Nov 2014
DOIs
Publication statusPublished - Mar 2015
Externally publishedYes

Bibliographical note

Publisher Copyright:
© 2014, Springer-Verlag Berlin Heidelberg.

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being

Fingerprint

Dive into the research topics of 'Roll and pitch set-up errors during volumetric modulated arc delivery: Can adapting gantry and collimator angles compensate?'. Together they form a unique fingerprint.

Cite this