Stray light in cone beam optical computed tomography: I. Measurement and reduction strategies with planar diffuse source

  • London Health Sciences
  • Western University
  • London Regional Cancer Program

Research output: Contribution to journalArticleAcademicpeer-review

11 Citations (Scopus)

Abstract

Optical cone-beam computed tomographic (CBCT) scanning of 3D radiochromic dosimeters may provide a practical method for 3D dose verification in radiation therapy. However, in cone-beam geometry stray light contaminates the projection images, degrading the accuracy of reconstructed linear attenuation coefficients. Stray light was measured using a beam pass aperture array (BPA) and structured illumination methods. The stray-to-primary ray ratio (SPR) along the central axis was found to be 0.24 for a 5% gelatin hydrogel, representative of radiochromic hydrogels. The scanner was modified by moving the spectral filter from the detector to the source, changing the light's spatial fluence pattern and lowering the acceptance angle by extending distance between the source and object. These modifications reduced the SPR significantly from 0.24 to 0.06. The accuracy of the reconstructed linear attenuation coefficients for uniform carbon black liquids was compared to independent spectrometer measurements. Reducing the stray light increased the range of accurate transmission readings. In order to evaluate scanner performance for the more challenging application to small field dosimetry, a carbon black finger gel phantom was prepared. Reconstructions of the phantom from CBCT and fan-beam CT scans were compared. The modified source resulted in improved agreement. Subtraction of residual stray light, measured with BPA or structured illumination from each projection further improved agreement. Structured illumination was superior to BPA for measuring stray light for the smaller 1.2 and 0.5 cm diameter phantom fingers. At the costs of doubling the scanner size and tripling the number of scans, CBCT reconstructions of low-scattering hydrogel dosimeters agreed with those of fan-beam CT scans.

Original languageEnglish
Pages (from-to)2893-2909
Number of pages17
JournalPhysics in Medicine and Biology
Volume61
Issue number7
DOIs
Publication statusPublished - 17 Mar 2016
Externally publishedYes

Bibliographical note

Publisher Copyright:
© 2016 Institute of Physics and Engineering in Medicine.

Fingerprint

Dive into the research topics of 'Stray light in cone beam optical computed tomography: I. Measurement and reduction strategies with planar diffuse source'. Together they form a unique fingerprint.

Cite this