Skip to main navigation Skip to search Skip to main content

Subharmonic, non-linear fundamental and ultraharmonic imaging of microbubble contrast at high frequencies

Research output: Contribution to journal › Article › Academic › peer-review

33 Citations (Scopus)

Abstract

There is increasing use of ultrasound contrast agent in high-frequency ultrasound imaging. However, conventional contrast detection methods perform poorly at high frequencies. We performed systematic in vitro comparisons of subharmonic, non-linear fundamental and ultraharmonic imaging for different depths and ultrasound contrast agent concentrations (Vevo 2100 system with MS250 probe and MicroMarker ultrasound contrast agent, VisualSonics, Toronto, ON, Canada). We investigated 4-, 6- and 10-cycle bursts at three power levels with the following pulse sequences: B-mode, amplitude modulation, pulse inversion and combined pulse inversion/amplitude modulation. The contrast-to-tissue (CTR) and contrast-to-artifact (CAR) ratios were calculated. At a depth of 8 mm, subharmonic pulse-inversion imaging performed the best (CTR = 26 dB, CAR = 18 dB) and at 16 mm, non-linear amplitude modulation imaging was the best contrast imaging method (CTR = 10 dB). Ultraharmonic imaging did not result in acceptable CTRs and CARs. The best candidates from the in vitro study were tested in vivo in chicken embryo and mouse models, and the results were in a good agreement with the in vitro findings. (C) 2015 World Federation for Ultrasound in Medicine & Biology.
Original languageUndefined/Unknown
Pages (from-to)486-497
Number of pages12
JournalUltrasound in Medicine and Biology
Volume41
Issue number2
DOIs
Publication statusPublished - 2015

Research programs

  • EMC COEUR-09

Cite this